Production Introduction



Containment Isolator


TAILIN containment isolator is designed for protecting the operators when weighing, dispensing, drying, pressing, filling highly potent and toxic products.


Experience:


TAILIN has been committed to the development of isolation technology and its application in the pharmaceutical industry, from sterility testing in the QC department, biosafety protection, and isolator technology for production (aseptic packaging, weighing, batching, crushing, sampling, etc. ) with RABS.


We can provide a variety of isolators to provide comprehensive technical solutions for product sterility protection or toxic containment. 


The newly introduced containment isolator (negative pressure isolator) is mainly used for production operations such as weighing, dispensing, and sampling of highly active and allergenic drugs.




Features:


  • OEB 4 or OEB5

  • H14 cylinder HEPA filters (PUSH-PUSH)

  • Continuous liner bag system for a material safety transfer

  • Siemens PLC control with HMI

  • Automatic control and operation

  • Real-time monitoring of the differential pressure, humidity, temperature

  • Critical process parameters can be stored and printed.

  • 21 CFR Part 11 Compliance




customized

customized

customized




Technical Parameters:


Power supply: 220V 50Hz

Chamber Material: SUS316L

Airflow model: Turbulent

Chamber Pressure: -80Pa to 50Pa adjustable

Air changes rate: More than 24 times/hour

Noise: <65dB(A) during negative pressure holding

Leak tightness: Meet ISO10648-2 requirements



Options:


Laminar flow with Bag-in-Bag-out H14 HEPA filters

Explosion-proof

The vibration-free platform for precision weighing

WiFi-controlled glove integrity tester (GIT-WLAN)

Wooden Mock-Up



Services:


Complete validation protocols to comply with FDA, EMA, GMP, 21CFR Part 11

FAT, SAT, Training